Objective: To determine the diagnostic performance of CT in the assessment of mild hepatic steatosis by comparison with MR mDIXON-Quant as a reference standard, and to explore their clinical applications. Methods: In this prospective study 169 volunteers were included. Each subject underwent CT and MR mDIX-ON-Quant examinations. Hepatic steatosis evaluations were performed via liver attenuation alone (CT L), liver to spleen attenuation ratio (CT L/S), difference between liver and spleen attenuation (CT L-S), and MR mDIX-ON-Quant imaging. The effectiveness of CT L, CT L/S, and CT L-S in diagnosing hepatic steatosis severity of ≥5%, ≥10%, and ≥15% was compared, using mDIX-ON-Quant results as standard. results: 65 subjects exhibited mild hepatic steatosis. Hepatic steatosis measurement with mDIXON-Quant was strongly correlated with the three CT methods. Using cutoff value, the sensitivity and specificity of diagnosing hepatic steatosis ≥5, ≥10, and ≥15% were 64.6, 91.3, 100%, and 90.4, 89.7, 93.0% for CT L; 50.8, 87.0, 100%, and 96.2, 98.6, 97.5% for CT L/S; and 67.7, 87.0, 100%, and 81.7, 98.6, 97.5% for CT L-S, respectively. ROC analysis indicated that 58.9, 56.5, and 52.8 HU for CT L; 1.06, 0.98, and 0.90 HU for CT L/S; and 6.21,-1.04, and -4.93 HU for CT L-S were cutoff values for diagnosing hepatic steatosis ≥5%,≥10%, and ≥15%, respectively. Conclusions: The three CT methods exhibit better agreements with mDIXON-Quant imaging for diagnosing hepatic steatosis ≥10%. Hence, CT and mDIXON-Quant could serve as suitable tools for the accurate quantification of mild hepatic steatosis. significant finds of the study: The close agreement between the three different CT methods (based on our cutoff values) and mDIXON-Quant imaging suggests that CT could accurately diagnose hepatic steatosis ≥10%. Thus, CT and mDIXON-Quant imaging can accurately measure mild hepatic steatosis. What this study adds: Only few studies have compared hepatic steatosis quantification between CT and mDIXON-Quant. We are the first to determine the diagnostic performance of unenhanced CT for quantitatively assessing mild hepatic steatosis, in reference to magnetic resonance mDIXON-Quant imaging. © 2018 The Authors. Published by the British Institute of Radiology.

Comparison of CT and magnetic resonance mDIXON-Quant sequence in the diagnosis of mild hepatic steatosis

Veronese, Nicola;
2018-01-01

Abstract

Objective: To determine the diagnostic performance of CT in the assessment of mild hepatic steatosis by comparison with MR mDIXON-Quant as a reference standard, and to explore their clinical applications. Methods: In this prospective study 169 volunteers were included. Each subject underwent CT and MR mDIX-ON-Quant examinations. Hepatic steatosis evaluations were performed via liver attenuation alone (CT L), liver to spleen attenuation ratio (CT L/S), difference between liver and spleen attenuation (CT L-S), and MR mDIX-ON-Quant imaging. The effectiveness of CT L, CT L/S, and CT L-S in diagnosing hepatic steatosis severity of ≥5%, ≥10%, and ≥15% was compared, using mDIX-ON-Quant results as standard. results: 65 subjects exhibited mild hepatic steatosis. Hepatic steatosis measurement with mDIXON-Quant was strongly correlated with the three CT methods. Using cutoff value, the sensitivity and specificity of diagnosing hepatic steatosis ≥5, ≥10, and ≥15% were 64.6, 91.3, 100%, and 90.4, 89.7, 93.0% for CT L; 50.8, 87.0, 100%, and 96.2, 98.6, 97.5% for CT L/S; and 67.7, 87.0, 100%, and 81.7, 98.6, 97.5% for CT L-S, respectively. ROC analysis indicated that 58.9, 56.5, and 52.8 HU for CT L; 1.06, 0.98, and 0.90 HU for CT L/S; and 6.21,-1.04, and -4.93 HU for CT L-S were cutoff values for diagnosing hepatic steatosis ≥5%,≥10%, and ≥15%, respectively. Conclusions: The three CT methods exhibit better agreements with mDIXON-Quant imaging for diagnosing hepatic steatosis ≥10%. Hence, CT and mDIXON-Quant could serve as suitable tools for the accurate quantification of mild hepatic steatosis. significant finds of the study: The close agreement between the three different CT methods (based on our cutoff values) and mDIXON-Quant imaging suggests that CT could accurately diagnose hepatic steatosis ≥10%. Thus, CT and mDIXON-Quant imaging can accurately measure mild hepatic steatosis. What this study adds: Only few studies have compared hepatic steatosis quantification between CT and mDIXON-Quant. We are the first to determine the diagnostic performance of unenhanced CT for quantitatively assessing mild hepatic steatosis, in reference to magnetic resonance mDIXON-Quant imaging. © 2018 The Authors. Published by the British Institute of Radiology.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14245/13192
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 29
social impact